Temperature range and temperature limit for pressure sensors ? is there an improvement? My intuitive answer will be: Yes! The initial term describes a section and the second its border. On second glance, however, I must conclude that both words ultimately express the same thing in relation to temperatures: Range and limit are defined by a lower and upper value, for instance 0 ? 100 �C. The relevant standard nevertheless defines an improvement. Why?
IEC 61987 speaks of two different specification characteristics
The standard referred to is IEC 61987. This deals, among other things, with the properties of fluid sensors, which likewise incorporate pressure sensors. With ?range? and ?limit?, the typical designates two different specification characteristics. Accordingly, the temperature range describes the span where the instrument specifications must apply ? first and foremost, the accuracy. The temperature limit, alternatively, indicates the min/max values between which the instrument could be operated without damage. With this particular, the instrument specifications do not have to be honored at all.
What may sound a bit pedantic, makes perfect sense from a technical perspective. This is often illustrated by the following example of a pressure sensor: The instrument is supposed to provide solid measured values at an ambient temperature range of 0 ? 100 �C. Concurrently, the sensor should never suffer any damage at ambient temperatures between -20 �C and 0 �C. In this range, however, it generally does not have to provide accurate measuring results, or even measure.
The difference between temperature range and temperature limit is plausible
This sounds paradoxical initially, but is plausible on closer inspection. Pressure sensor elements, i.e. the actual measuring components, exhibit a relatively large, often non-linear temperature error. Without further measures, a trusted pressure measurement will be impossible. Therefore, the maker must compensate for the temperature so as to bring the error down to an acceptable level. From an economic perspective, the limitation to a selected temperature range is practical, or is even absolutely necessary.
The distinction between temperature range and temperature limit applies to both ambient temperature and the medium temperature. Additionally it is useful for other specification characteristics, for instance overpressure.
Conclusion
Yes, there is a difference between range and limit in the normative world of pressure sensor technology. And yes, Pus makes technical sense. However, I doubt whether the normal user, without understanding of standards, understands it intuitively. Which inevitably results in the question of whether there is a better linguistic distinction. But, I must admit, the solution is outside my ?range?.
Note
More info on our pressure sensors are available on the WIKA website. Do you wish to buy pressure sensors? You will discover some of our standard designs in our WIKA online-shop. Should you have any questions, your contact will gladly help you.
Also read our posts
What does temperature compensation or compensated temperature range for pressure sensors mean?
Active and passive temperature compensation of pressure sensors
Temperature coefficients (TC) of pressure sensors

Leave a Reply