Temperature coefficients (TC) of pressure sensors

Since temperature affects the measuring accuracy of a pressure sensor, there always remains a small temperature error in the rated temperature range despite a wide range of compensation measures. This error is often expressed in the data sheets of manufacturers of pressure sensors as temperature coefficient (abbr. TC). This coefficient describes a (linear) error, beginning with a reference point, which in most cases is room temperature.
Accordingly, the temperature error at room temperature is zero and increases with increasing difference of the temperature from room temperature with the specified coefficient in linear fashion (see figure). Basically, not merely the zero-point temperature coefficient (abbr. TC0), but also that of the span (TCC) should be considered individually. No Good -point error describes the effect of temperature on the zero signal. The error of the span specifies the result of temperature on the entire scale value. The individual temperature coefficients of zero point and of span are usually specified as magnitudes, and therefore they may be either positive or negative.
If within an individual instrument the zero-point error gets the same sign because the error of the span, both of these temperature errors may add up in worst case. An average value for the common zero-point temperature coefficient of a pressure sensor is: 0.2 % / 10 K. Connect to WIKA pressure sensors.